2 troubleshooting temperature problems -10, Table 10-3 – Teledyne 300M User Manual

Page 106

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 10-3: V/F Board Jumpers - Factory Settings

Factory Set Jumpers

Jumper

Setting

JP1 1-2

JP2 1-2

B12 3-4

B14 ON

B15

Set to match input

line frequency

Table 10-4: V/F Board Switch Settings - Ranges for Analog Output

User Set Switches

Switch

100 mV

Full Scale

1 V

Full Scale

5 V

Full Scale

10 V

Full Scale

S1 (Recorder Output)

1, 6

1, 5

1, 4

1, 3

S2 (DAS Output)

1, 6

1, 5

1, 4

1, 3

S3 (Test Output

1, 6, 7

1, 5, 7

1, 4, 7

1, 3, 7

S4 (Spare)

1, 6, 7

1, 5, 7

1, 4, 7

1, 3, 7

3. If voltages and jumper settings are correct, the V/F card is faulty and should be replaced.

The primary functions of the board can be divided into three areas:

1. Channels of Multiplexed Analog input to an Analog to Digital converter

2. Independent Digital to Analog converters

3. Digital I/O Lines configured as 24 outputs and 8 inputs

The following sections describe each of these functional areas.