Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 4-1: Setup Functions (Continued)

Button Sequence

Function

Default

Limits

SETUP-MORE-DIAG-
SIGNAL I/O

View the state of
internal signals

N/A N/A

SETUP-MORE-DIAG-
ANALOG OUTPUT

Generate Analog
Output Test Pattern

N/A N/A

SETUP-MORE-DIAG-
D/A CALIBRATION

Calibrate D/A and A/D
Converters

N/A N/A

SETUP-MORE-DIAG-
ELECTRICAL TEST

Generate Electrical
Test Output

N/A N/A

SETUP-MORE-DIAG-
DARK
CALIBRATION

Adjust Dark Offset

125mV

75 - 175 mV

SETUP-MORE-DIAG-
FLOW
CALIBRATION

Calibrate sample
flowrate

800 cc/m

800 cc/m

SETUP-MORE-
DIAG- TEST CHAN
OUTPUT

Select TEST to Analog
Output

None

SETUP-MORE-DIAG-
RS-232 OUTPUT

Generate RS-232
Output Test Pattern

N/A N/A

SETUP-MORE-
ALRM-ALM1

Enable alarm 1 and set
limit

ON 100

PPM

SETUP-MORE-
ALRM-ALM2

Enable alarm 2 and set
limit

ON 300

PPM

4-9