Table 5-2: I/O Signals (Continued)

#

Signal

Control

Description

32

DCPS_VOLTAGE

NO

DC power supply composite voltage output. Typically
2500 mV.

33

DAC_CHAN_0

NO

Output of DAC 0(REC) in mV

34

DAC_CHAN_1

NO

Output of DAC 1(DAS) in mV

35

DAC_CHAN_2

NO

Output of DAC 2(TEST) in mV

36

DAC_CHAN_3

NO

Output of DAC 3(Spare) in mV

37

CONC_OUT_1

YES

CO Reading (REC) in mV

38

CONC_OUT_2

YES

CO Reading (DAS) in mV

39

TEST_OUTPUT

YES

Test Channel in mV

40

SAMPLE_LED

YES

Sample LED on/off

41

CAL_LED

YES

Calibration LED on/off

42

FAULT_LED

YES

Fault LED on/off

5.2.2 Analog Output Test

This test cycles the analog output channels from 0% to 100% of full scale in 20% full scale
steps. It starts by outputting 0 volts to all channels and displaying a 0% button. Then, every five
seconds, the output is increased 20% FS and the button is changed accordingly. Thus, the button
(and the analog outputs) will cycle through the following value.

0%, 20%, 40%, 60%, 80%, 100%, 0%, ...

To pause the output at the current voltage, press the n% button. To resume automatic cycling,
press the n% button again.

5.2.3 Electric Test

This test activates a diagnostic circuit located on the Synchronous Demodulator board which
generates an artificial signal which simulates the output of the IR detector. This signal is injected
in place of the detector output.

When activated, Electric Test will automatically switch the analyzer into a 5000 PPM range and
result in the analyzer producing a constant, stable output (i.e. CO reading) of about 2500 PPM.
This test is particularly useful in isolating problems, since it exercises essentially all electronic
sub-systems of the analyzer but does not depend on the proper function of optical or pneumatic
subsystems.