Teledyne 300M User Manual

Page 44

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Example 4: to perform zero check once per day at 10:30 PM and a span check once per week
starting at 11:30 PM, 12/20/93.

1. Select any one of SEQx and program as example 1.

2. Select any other SEQx and program as example.

3. Avoid setting two or more sequences at the same time of the day.

3.5 Summary of Front Panel Check and Calibration

Controls

The calibration controls are summarized below in terms of the button sequences used to access
them.

Table 3-1: Calibration Controls

Button Sequence

Function

Default

Limits

CALZ

Begin zero check

-

-

CALZ-ZERO-ENTR

Adjust CO conc. to zero

-

-

CALZ-EXIT

Exit zero check

-

-

CALS

Begin span check

-

-

CALS-CONC-ENTR

Expected CO span value

400 PPM

1 - 20000 PPM

CALS-SPAN-ENTR

Adj. CO conc. to span val.

-

-

CALS-EXIT

Exit span check

-

-

CAL

Begin M-P cal.

-

-

CAL-ZERO-ENTR

Adj. CO conc. zero value

-

-

CAL-CONC-ENTR

Expected CO span value

400 PPM

1 - 20000 PPM

CAL-SPAN-ENTR

Adj. CO conc. to span val.

-

-

CAL-EXIT

Exit M-P cal.

-

-

SETUP-ACAL-SEQ1

Setup auto-cal SEQ1

Disabled

SETUP-ACAL-SEQ2

Setup auto-cal SEQ2

Disabled

SETUP-ACAL-SEQ3

Setup auto-cal SEQ3

Disabled

3-8

Advertising