Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 4-1: Setup Functions

Button Sequence

Function

Default

Limits

SETUP-CFG-NEXT List

Software

Configuration

CO Mach

N/A

SETUP-ACAL-MODE Define/Change

AutoCal Sequences

Disabled

Zero, Span, Zero-
Span

SETUP-DAS-EDIT Define/Change

DAS

Data Channels

SETUP-DAS-VIEW

View DAS Data

N/A

N/A

SETUP-RNGE-MODE

Set Range Mode

Single

Single, Dual, Auto

SETUP-RNGE-SET

Set D/A output range

500 PPM

1-20,000 PPM

SETUP-RNGE-UNIT

Set Measurement Units

PPM

PPM. mg/m

3

, ug/m

3

SETUP-PASS Password

Enable

ON OFF-ON

SETUP-CLK-TIME Set

Time-of-Day

00:00

00:00-23:59

SETUP-CLK-DATE

Set Current Date

01 JAN 00

31 DEC 99

SETUP-MORE-
COMM-BAUD

RS-232 baud rate

2400 baud

300, 1200, 2400,
4800, 9600, 19.2

SETUP-MORE-
COMM-ID

Analyzer ID number

0000

0000-9999

SETUP-MORE-
VARS-
DAS_HOLD_OFF

Set Hold-Off Interval

15 min

1 - 60 min

SETUP-MORE-
VARS- DYN_ZERO

Enable Remote
Dynamic Zero
Adjustment

OFF OFF-ON

SETUP-MORE-
VARS- DYN_SPAN

Enable Remote
Dynamic Span
Adjustment

OFF OFF-ON

SETUP-MORE-
VARS- RS232_MODE

Set RS-232 Mode

8

SETUP-MORE-
VARS- CLOCK_ADJ

Set Clock Adjustment
Rate

0

-15 to +15

4-8