Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 5-2: I/O Signals (Continued)

#

Signal

Control

Description

19

ST_HIGH_RANGE

YES

Status Bit - Autorange High Range

Logic High = M300M in high range

Logic Low = M300M in low range

20

ST_SYSTEM_OK

YES

Status Bit - System OK

Logic High = No instrument warning present

Logic Low = 1 or more alarm present

21

ST_BENCH_ALARM

YES

Status Bit - Bench Temperature Alarm

Logic High = Bench Temp out of spec

Logic Low = Bench Temp in spec

22 ST_SOURCE_ALAR

M

YES

Status Bit - IR Source Alarm

Logic High = IR Source output too low

Logic Low = IR Source output normal

23

ST_WHEEL_ALARM

YES

Status Bit - Wheel Temperature Alarm

Logic High = Wheel Temp out of spec

Logic Low = Wheel Temp in spec

24

CO_MEASURE

NO

IR detector reading during measure phase. Typically
2500 - 4500 mV.

25

CO_REFERENCE

NO

IR detector reading during reference phase. Typically
2500 - 4500 mV.

26

SAMPLE_PRESURE

NO

Sample pressure in mV. Typical sea level value =
4300 mV for 29.9" HG-A.

27 VACUUM_PRESSUR

E

NO

Sample flow in mV

28

SAMPLE_TEMP

NO

Sample temp in mV

29 BENCH_TEMP

NO

Optical Bench temp. typically 2270 mV for 48

°C

30 WHEEL_TEMP

NO

Filter Wheel temp. typically 4770 mV for 68

°C

31

BOX_TEMP

NO

Internal analyzer temp. in mV

(table continued)