Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 2-3: Test Measurements

Test Message

Meaning

TIME=xx:xx:xx

Current time-of-day (HH:MM:SS)

RANGE=xxxx PPM

Analog output full-scale range

STABIL=x.xxx PPM

Standard Deviation of CO readings

CO MEAS=xxxxx MV

Current V/F measure channel (mV)

CO REF=xxxxx MV

Current V/F reference channel (mV)

MR RATIO=X.XXX

Ratio of the Reference and Measure values

AZERO OFFS=x.xxxxx

Autozero offset

PRES=xxx IN-HG-A

Absolute sample pressure – inches Hg

VAC = xxx IN-HG-A

Absolute vacuum pressure – inches Hg

SAMPLE FL=xxx CC/M

Sample flow through analyzer (cc/min)

SAMPLE TEMP=xxx C

Temperature in the absorption cell

BENCH TEP=xxx C

Optical Bench Temperature

WHEEL TEP=xxx C

Filter Wheel Temperature

BOX TEMP=xxx C

Internal box temperature (degrees C)

DCPS=xxxxxx MV

DC power supply (mV)

SLOPE=x.xxx

Internal formula - Slope

OFFSET=xx.x MV

Internal formula - Offset