10.6.6 Checking the Opto Interrupter

Correct operation of the Opto Interrupter on the gas filter wheel can be confirmed by connecting
an oscilloscope U6, Pin 11 on the Sync Demodulator board and comparing the waveform to
Figure 10-1. The waveform should be symmetrical and 5 Volts peak to peak.

Figure 10-1: Opto Pickup Waveform

Figure 10-2: Detector Waveform