Programmable push buttons, Able, 3: t – Teledyne 300M User Manual

Page 33: Easurements, 2 programmable push buttons

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 2-4: Warning Messages

Warning Message

Meaning

SYSTEM RESET

Issued whenever analyzer is powered on

RAM INITIALIZED

RAM was erased (incl. DAS reports)

SOURCE WARNING

IR source < 2500 OR >= 5000 mV

AZERO WARNING

SAMPLE FLOW WARN

Sample flow < 500 cc/m or > 1200 cc/m

SAMPLE PRESS WARN

Sample pressure < 15 or > 35 In-Hg-A

SAMPLE TEMP WARN

Sample temperature < 10

°C or > 50°C

BOX TEMP WARNING

Box temp. < 12

°C or > 48°C

BENCH TEMP WARNING

Optical Bench < 43

°C or > 53°C

WHEEL TEMP WARNING

Filter Wheel < 63

°C or > 73°C

CANNOT DYN ZERO

CO offset < -1500 or > +1500 mV

CANNOT DYN SPAN

CO slope < 0.5 or > 2.0

V/F NOT DETECTED

A/D - I/O card not installed or bad

SYNC WARNING

No modulation on detector output

DCPS WARNING

The menu field changes depending on the mode of the analyzer and the buttons that have been
pressed. It indicates the current function of each of the 8 push buttons below the display. See
Section 2.2.2 for information on using the push buttons.

2.2.2 Programmable Push Buttons

The 8 push buttons below the display are programmable by the CPU and their functions change
depending on the mode of the analyzer or the operations being performed. The legend above a
button identifies its current function. If there is no legend above a button, it has no function and
will be ignored if pressed.

2-7

Advertising