Teledyne 300M User Manual

Page 43

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Examples of possible sequences are as following under any one of three available SEQx.

Example 1: To perform a 15 minute zero check once per day at 10:30 PM, 12/20/93.

1. MODE: ZERO

2. TIMER ENABLED: ON

3. STARTING DATE: 12/20/93

4. STARTING TIME: 22:30

5. DELTA DAYS: 1

6. DELTA TIME: 00:00

7. DURATION: 15

8. CALIBRATE: OFF

Example 2: To perform a 15 min zero span check once per day retarding 15 minutes everyday
starting at 11:30 PM, 12/20/93.

1. MODE: ZERO-SPAN

2. TIMER ENABLED: ON

3. STARTING DATE: 12/20/93

4. STARTING TIME: 23:30

5. DELTA DAYS: 0

6. DELTA TIME: 23:45

7. DURATION: 15

8. CALIBRATE: OFF

Example 3: To perform span check once per week starting at 11:30 PM, 12/20/93.

1. MODE: SPAN

2. TIMER ENABLED: ON

3. STARTING DATE: 12/20/93

4. STARTING TIME: 23:30

5. DELTA DAYS: 7

6. DELTA TIME: 00:00

7. DURATION: 15

8. CALIBRATE: OFF

3-7

Advertising