Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 10-1: Warning Messages

Warning Message

Meaning

Corrective Action

SOURCE WARNING

The CO REF value is greater
than 5000 mV or less than
2500 mV

Check and adjust the Sync Demodulator
and optical alignment as described in
Section 10.6.6 and 10.6.5

SYNC ERROR

No modulation is present on the
output of the IR detector.

Check IR source, IR detector/pre-amp,
and Opto interrupter

BENCH HEAT
SHUTDOWN

Temperature control of the
Optical bench cannot be
maintained at its 48°C set point

Check Optical Bench heater and
thermistor

SAMPLE PRESSURE
WARNING

The Sample Pressure is less
than 15"Hg or is greater than
35"Hg

Check for pressure transducer problems
as described in Section 10.6.5 and
Section 10.6.1

SAMPLE FLOW
WARNING

The sample flow is less than
500 cc/min or greater than 1200
cc/min.

Check for pneumatic system problems as
described in Section 10.6.1. Check for
flow transducer problems as described in
Section 10.6.5

BOX TEMP
WARNING

The inside chassis temp is less
than 10°C or is greater than
50°C

See Section 10.6.2

SAMPLE TEMP
WARNING

The Sample Temperature is less
than 10

°C or is greater than

50°C

See Section 10.6.2

CANNOT DYN ZERO

An offset of more than ±5 PPM
would be required to Zero
adjust the analyzer

See Section 10.5.5

CANNOT DYN SPAN

A slope of less than 0.5 or
greater than 2.0 would be
required to Span adjust the
analyzer

See Section 10.5.6

V/F NOT INSTALLED

The CPU is unable to
communicate with the V/F
Board

Check and re-seat CPU and V/F board.
See Section 10.6.3

SYSTEM RESET

A power Off-On cycle has
occurred

None required

RAM INITIALIZED

Dynamic memory has been re-
initialized in response to the
installation of a new PROM or
memory chip

None required