Tatus, Ontrol, Able – Teledyne 300M User Manual

Page 73: 4: s, Eports

Advertising
background image

Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

Table 7-5: Control Commands

Command Message

Meaning

"C ZERO<CRLF>"

Do a zero check

"C COMPUTE ZERO<CRLF>"

Calibrate Zero point

1

"C SPAN<CRLF>"

Do a span check

"C COMPUTE SPAN<CRLF>"

Calibrate Span point

1

"C ASEQ1<CRLF>"

Do a auto-cal sequence

2

"C ASEQ2<CRLF>"

Do a auto-cal sequence

2

"C ASEQ3<CRLF>"

Do a auto-cal sequence

2

"C ABORT<CRLF>"

Aborts auto-cal sequence

"C EXITZ<CRLF>"

Exit zero cal only

"C EXITS<CRLF>"

Exit span cal only

"C EXIT<CRLF>"

Exit zero span or hold

1

Executed only if the instrument is in the proper calibration mode and concentration is within

calibration limits. This command adjusts slope and offset values.

2

Initiated only If automatic calibration sequence setup is programmed and enabled.

NOTE

The commands in Table 7-5 can only be entered via the RS-232 port

when the analyzer is in the sample mode.

When a control command is issued, the CPU will respond by issuing a status report. For example
if the host computer issues the command.

C CALZ<CRLF>

to do a zero check, the CPU will send the status report

C DDD:HH:MM IIII START ZERO CALIBRATION<CRLF>

to the RS-232 output.

7-7

Advertising