Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A

7.2 Warnings

Whenever a warning message is displayed on the display, it is also sent to the RS-232 output.
See Table 2-4 for a list of the warning messages. These messages are very helpful when trying to
track down a system problem and for determining whether or not DAS average data is actually
valid. The message format is:

W DDD:HH:MM IIII WARNING MESSAGE<CRLF>

An example of an actual warning message is:

W 194:11:03 0000 SAMPLE FLOW WARN<CRLF>

Warnings may be cleared via the RS-232 interface by issuing a command of the form:

W COMMAND<CRLF>

Where "COMMAND" indicates which warning message to clear. For example, to clear the
"SAMPLE FLOW WARN" message, the host computer can issue the command:

W WSMPFLOW<CRLF>

Attempting to clear a warning which is not active has no effect. The Table 7-3 lists the command
to use to clear each possible warning message.