KEW6305

8.8

● Displayed items

<Screen 1, Upper row:

Target demand value

(W) >

It is set at “Setting 15”. Set the desired value.

<Screen 1, Middle row:

Predicted demand value

(W) >

Predicted values of average electrical power (demand values) which will be after the elapse of
demand measurement interval with present loads are displayed right after starting a measurement.
The predicted values are calculated successively when loads vary.


<Screen 1, Lower row & Screen 2, Lower row:

Present demand value

(W) >

This is the Present Demand Value(see definitions)

<Screen 2, Upper row:

Remaining time

>

This is the count down, in 1 sec. decrements to the end of the interval.

<Screen 2, Middle row:

Load factor

(%) >

This is the load factor(See definitions)

<Screen 3, Upper row & Middle row:

Date and time

>

These are the time and when the max. demand recorded so far from the start of the survey was
measured.

<Screen 3, Lower row:

Max. demand value

(W) >

The max. demand value measured between the beginning and the end of
measurement is displayed.