8.12

8-6 Saving data

When integration or demand measurement starts, the measured data will be saved automatically.
There are two locations where data can be saved.

* SD card

: Max. 511 files can be saved.

* Internal memory

: Max. 4 files can be saved.

Data is saved to a SD card automatically when a SD card has been inserted before turning on the
instrument. If the SD card has not been inserted, data is saved automatically to the internal memory

8.6.1 Saving procedure

* When a survey is started (manually or automatically) a file is opened
* Data is saved at the end of each integration interval (“Setting 09”).

Elapsed time

Demand interval

1 2 3

n-1

n

Start of demand Data saving point End of demand

(manually or preset time & date)

(manually or preset time & date)

* When the survey is closed (manually or automatically) the file is closed

* All the recorded parameters at each data saving point are saved to one file.

NOTE

* Never set the Function switch to OFF position during a survey otherwise the measured data may be lost.

* In the following cases, the file number becomes “001”:

(1) when the file number has exceeded 999;
(2) after system reset

* If the same file no. exists, the old one will be overwritten.