KEW6305

8.10

● To start measurement automatically at the preset time and date

1

Set the start time and date on

SET UP

range.

2

Then set the Function switch to

DEMAND

range, and press the

START/STOP

key.

3

File number screen is displayed for about 1 sec., followed by the measurement screen. The

instrument goes into stand-by mode for measurement. LED status indicator while the instrument is
in stand-by mode.

File number screen

(displayed for about 1 sec.)





Measurement screen



4

Measurement starts at the preset time and date, and the LED status indicator stop flickering and
remain on for the duration of the survey.







NOTE

* The start time and date should be set after the current time in such a way to give enough time to the

user to complete all settings before the survey starts.

* When the start time and date are set before the current time, measurement starts immediately upon

pressing the

START/STOP

key.

* When the preset start time and date come after the preset stop time and date, measurement stops right

after the start of measurement.

* Even if the start and stop time have been preset and the instrument is in standby mode, pressing the

START/STOP

key at least 2 sec. will release

the stand-by mode and start a survey in Manual mode. This renders the start/ stop time and date

settings ineffective.