KYORITSU 6305 Instruction Manual User Manual

Page 46

Advertising
background image

4.19

KEW6305

Setting 17” Demand warning cycle

The buzzer will sound when a predicted demand value exceeds a target demand value during
demand measurement.
For further details, please refer to “Section 8”: Demand measurement.
According to the demand measurement interval, which has been set at Setting 16, warning cycle can
be set to as follows.

Demand measurement cycle

“Setting 16”

Warning cycle

10 / 15 min.

1 / 2 / 5 min.

30 min.

1 / 2 / 5 / 10 / 15
min.

Default value (or after system reset)

10 min.

1 Use the

Cursor

key on the selection screen, and select “Setting 17”.

2 Press the

ENTER

key to get the instrument into setting change mode.

3 Previous setting (or default value: 10 min.) blinks. Press the

Cursor

key to select any desired time,

and then press the

ENTER

key after making necessary change.