KEW6305

4.6

Setting 04” Current range

Selectable Current range differs by clamp sensor selected at “Setting 03”.

Clamp sensor (“Setting 03”)

Current range

50A (M-8128)

1 / 5 / 10 / 25 / 50A / AUTO

100A (M-8127)

2 / 10 / 20 / 50 / 100A / AUTO

200A (M-8126)

4 / 20 / 40 / 100 / 200A / AUTO

500A (M-8125)

10 / 50 / 100 / 250 / 500A / AUTO

1000A (M-8124)

20 / 100 / 200 / 500 / 1000A / AUTO

3000A

300 / 1000 / 3000A

Default value (or after system reset)

AUTO

* Selecting “AUTO” activates auto-ranging function and the measuring range will be automatically

switched between the lowest and highest ranges.

1 Use the

Cursor

key on the selection screen, and select “Setting 04”.

2 Press the

ENTER

key to get the instrument into setting change mode.

3 Present setting (or default value: AUTO) blinks. Select the appropriate current range with the

Cursor

key, and then press the

ENTER

key after making necessary change.

NOTE:

* When a type of clamp sensor (“Setting 04”) is changed, current range may be changed to the

corresponding range automatically.

* Accurate results may not be obtained if the Clamp sensors in use do not match the setting done for

sensor.

* Using the auto-ranging function can measure wide range of input signals, however, accurate results

may not be obtained when measuring loads that flactuate so widely within 1 sec..