7.9

KEW6305

NOTE

* Never set the Function switch to OFF position during a survey otherwise the survey may be lost.

* The file number becomes “001” when;

(1) when the file number has exceeded 999
(2) after system reset

* If the same file no. exists, the old one will be overwritten.

7.5.2 Limitations of saving

● Limitation of saving (before starting a survey)

In the following cases, a survey cannot be started (manually or automatically) by pressing the

START/STOP

key.

< In case data is saved to SD card >

* When 511 files have been saved to the SD card;

mark appears and further data cannot be

saved.

Some files can to be deleted via the PC, otherwise all the saved files in the SD card can be deleted
using “

Setting 19” of Section 4 in this manual.

< In case the data is saved to internal memory >

* When 4 files have been saved to the internal memory;

mark appears and further data

cannot be saved.

● Limitations of saving (during a survey)

When the capacity of SD card or internal memory has been exceeded during a survey; measurement
continues but

mark appears on the display screen and further data will not be saved.

Press the

START/STOP

key 2 sec. or more and stop the survey once.

Refer to preceding page and delete the unnecessary file.

NOTE

* For further details about the capacity of the CF card and internal memory, refer to “

Section 9: CF card/

Internal memory” in this manual.