KYORITSU 6305 Instruction Manual User Manual

Page 31

Advertising
background image

KEW6305

4.4

Setting 02” Voltage rang

Selecting a measurement range so that the estimated inputs will be close to the full scale value is

recommended to obtain accurate results. Recommended range selections are: 150V range for rated

voltages between 100 - 120V, 300V range for 200 - 240V and 600V range for 400 – 440V.

Setting item

150V / 300V / 600V

Default value (or after system reset)

300V

1 Use the

Cursor

key on the selection screen, and select “Setting 02”.

2 Press the

ENTER

key to get the instrument into setting change mode.

3 Present setting (or default value: 300V) blinks. Select the appropriate voltage range with the

Cursor

key, and then press the

ENTER

key after making necessary change.