2 inst measurement, 2 inst measurement ( w range), 1_ p p – KYORITSU 6305 Instruction Manual User Manual

Page 131

Advertising
background image

KEW6305

13.2


13.2 Inst measurement (

W

Range)

(1) Voltage V(i) [V]

Range

150/ 300/ 600V

Displayed digit

4 digits

Allowable input

10 to 110% of each range

Display range

5 to 130% of each range

Crest factor

2.5 or less

Accuracy

±0.2%rdg±0.2%f.s. (sine wave, 45 - 65Hz)

Input impedance

approx. 8.3MΩ

(2) Current A(i) [A]

Range

50A type (8128)

:1/ 5/ 10/ 25/ 50A/ AUTO

100A type(8127)

:2/ 10/ 20/ 50/ 100A/ AUTO

200A type(8126)

:4/ 20/ 40/ 100/ 200A/ AUTO

500A type(8125)

:10/ 50/ 100/ 250/ 500A/ AUTO

1000A type(8124)

:50/ 100/ 200/ 500/ 1000A/ AUTO

3000A type(8129)

:300/ 1000/ 3000A

Displayed digit

4 digits

Allowable input

10 to 110% of each range

Display range

1 to 130% of each range

Crest factor

3.0 or less (max1.4Vpeak)

Accuracy

±0.2%rdg±0.2%f.s. + accuracy of clamp sensor (sine wave, 45 - 65Hz)
* +1% f.s. at the lowest range.

Input impedance

approx. 100kΩ

(3) Active power P(i) [W]

Range

(Voltage range) x (Current range)

Displayed digit

4 digits

Accuracy ±0.3%rdg±0.2%f.s.

+ accuracy of clamp sensor (power factor 1, sine wave, 45 - 65Hz)

* +1% f.s. when the lowest current ranges is selected.

Effect of power
factor

±1.0%rdg
(indicated value when power factor 0.5 against power factor 1)

Polarity indication

Consumption : +( no mark), Regeneration: -

×1

1

_

P

P

=

×2

2

1 P

P

P

+

=

1P2W

×3

3

2

1

P

P

P

P

+

+

=

1P3W

2

1 P

P

P

+

=

3P3W

2

1 P

P

P

+

=

3P3W3A

3

2

1

P

P

P

P

+

+

=

Equation

3P4W

3

2

1

P

P

P

P

+

+

=

(4) Frequency f [Hz]

Accuracy ±3dgt
Displayed digit

3 digits

Allowable input

10 - 110% of V range (sine wave, 45 - 65Hz)

Display range

40.0 - 70.0Hz

Input source

V1

Advertising