For details about demand target value, please refer to “Section 8”: Demand measurement. Target
value can be selected between 0.1W and 999.9GW.

Value Unit

Demand target value

0.1 - 999.9

(can be set by 0.1)

W / kW / MW / GW

VA / k VA / M VA / G VA

Default value (or after

system reset)

100.0kW

1 Use the

Cursor

key on the selection screen, and select “Setting 15”.

2 Press the

ENTER

key to get the instrument into setting change mode.

3 Present setting (or default value: 100.0kW) blinks. Change the value and unit with the

Cursor

key.

4 Press the

ENTER

key after making necessary change.

Function of

Cursor

keys:

To select the digit or unit parameter subject to change.

To change the value of selected digit and unit parameter.

Either “W” or “VA” can be set as a unit.
The instrument can display and record the demand values of active and apparent power by switching
above unit.


NOTE:

* When the target value is set to 0.0, it is forcefully changed into 100.0.