Dual range – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 19

Advertising
background image

17

6.2.

Dual Range

Weighing range

GH-252 GH-202

Available minimum display

Precision range 0 g to 101 g 0 g to 51 g 0.01 mg 0.1 mg

1 mg

Standard range 101 g to 250 g 51 g to 220 g

0.1 mg

1 mg


The GH-252 and GH-202 are

equipped with two ranges "precision
range" and "standard range".

When weighing is started by

pressing the

ON:OFF

key, the

minimum display will be 0.1 mg.

Pressing the

RANGE

key will

switch these ranges alternately.



When a sample is weighed using the

precision range with a minimum display of
0.01 mg and the weight value exceeds the
precision range value, the minimum display
changes to 0.1 mg of the standard range.
When removing the sample, the minimum
display changes to 0.01 mg of the precision
range automatically.

When a tare weight (container mass value)

exceeds the precision range value, even if
the sample is within the precision range
value, 0.01 mg of the precision range can
not be selected for the minimum display. In
order to select 0.01 mg of the precision
range, remove the tare weight and press
the

RE-ZERO

key to cancel it.





When a minimum display of 0.1 mg or 1 mg is

selected by the

RANGE

key, the minimum

display is maintained even if the range is
changed during weighing.

Turn on the balance

Turn on the balance

Minimum display

0.1mg

Minimum display

0.01mg

Minimum display

1mg

Precision Range

Minimum display

0.1mg

Minimum display

1mg

Standard Range

Minimum display

0.01mg

Minimum display

0.1mg

Minimum display

0.01mg

Minimum display

0.01mg

Minimum display

0.1mg

Zero display after

removing the container

Minimum display 0.01mg

Zero display after

placing a container

Minimum display 0.1mg