A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 37

Advertising
background image

35

Class Item

and

Parameter

Description

0

Every measurement

1

2 seconds

2

5 seconds

3

10 seconds

4

30 seconds

5

1 minute

6

2 minute

7

5 minute

int
Interval time

8

10 minute

Interval time in the interval memory

mode when using prt 3, data 1

0

No output

d-no
Data number output

1

Output

Refer to "12. Data Memory"

0

No output

1

Time only

2

Date only

5-td
Time/Date output

3

Time and date

Selects whether or not the time

or date is added to the weighting

data. Refer to "10.7. Clock and

Calendar Function" for details.

0

No output

5-id
ID number output

1

Output

Selects whether or not the ID

number is output.

0

No pause

pU5e
Data output pause

1

Pause (1.6 seconds)

Selects the data output interval.

0

Not used

at-f
Auto feed

1

Used

Selects whether or not automatic

feed is performed.

0

No output

1

AD-8121 format

info
GLP output

2

General data format

Selects GLP output method.

Refer to "11. ID Number And GLP

Report "

for details.

0

Not used

dout
Data output

ar-d
Zero after output

1

Used

Adjusts zero automatically after data

is output

0

600

bps

1

1200

bps

2

2400

bps

3

4800

bps

4

9600

bps

bp5
Baud rate

5

19200 bps

0

7 bits, even

1

7 bits, odd

btpr
Data bit, parity bit

2

8 bits, none

0

CR LF

Crlf
Terminator

1

CR

CR: ASCII code 0Dh

LF: ASCII code 0Ah

0

A&D standard format

1

DP format

2

KF format

3

MT format

4

NU format

5if
Serial
interface

type
Data format

5

CSV format

Refer to "10.5. Description of Item

"Data Format".

: Factory settings.

Caution

The balance may not transmit the data completely at the specified refresh rate, depending

on the baud rate or data added to the weighing data such as time, date and ID number.

Advertising