Calibration test report using an internal mass

Note Calibration test does not perform calibration.

Key operation
1 Press and hold the

CAL

key to display

CC in

and release the key.

2

CC

is displayed and the balance is tested automatically.

3 The zero point is measured and the weight value is displayed for a few seconds.
4 Internal mass is weighed and the weight value is displayed for a few seconds.
5 If GLP output is used,

glp

is displayed and the calibration test report is output.

6 The balance returns to weighing mode after this test.


Commad

This calibration test report can be performed with commad TST.



Setting of "info 1"

Setting of "info 2"

AD-8121 printer format

General format



















Space, ASCII 20h

<TERM>

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration test type

Zero point value

Target weight value

Target weight

Signature

-----------

A

-

&

-

D

MODEL

-----

GH-300

S/N

-----

01234567

ID

------

LAB-0123

DATE

--

2004/07/01

TIME

----

12:34:56

CAL.TEST(INT.)
ACTUAL

-------

0.0000

--

g

----

+200.0002

--

g

TARGET

----

+200.0000

--

g

SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-