Csv format 5if type 5, Id number dout 5-id 1, Data number dout d-no 1 – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 43: Date dout 5-td 2 or 3, Time dout 5-td 1 or 3

Advertising
background image

41

L A B - 0 1 2 3

CR LF

2 0 0 4 / 0 7 / 0 1

CR LF

CR LF

1 2 : 3 4 : 5 6

CSV format

5if type 5

This format separates the data of A&D standard format and the unit by a comma ( , ).
This format outputs the unit even when the data is overloaded.
When a comma ( , ) is selected for the decimal point, the separators are set to semicolon ( ; ).
When the ID number, data number, time and date are added at "Data output (

dout

)" of the

function table, outputs ID number, data number, date, time and weighing data in this order
and separates each item by a comma and treats all the items as one group of data.







ID number

dout 5-id 1

The number to identify a specific balance.

This format consists of eight characters excluding the terminator.

Data number

dout d-no 1

This format outputs the data number just before the data is transmitted using the RS-232C
interface.

This format consists of six characters excluding the terminator.
When CSV format (

5if type 5

) is selected, the period ( . ) is replaced with a comma ( , ).


Date

dout 5-td 2

or 3

The date output order can be changed in " Time/Date output (5-td)" and "Clock (

Cl adj

)".

The year is output in a four-digit format.

Time

dout 5-td 1

or 3

This format outputs time in 24-hour format.



Data number

Terminator

LAB-0123, No,012, 2004/07/01, 12:34:56, ST,+0000.1278,~~g<CR><LF>

CR LF

S T , + 0 0 0 . 1 2 7 8 ,

g

CR LF

O L , + 9 9 9 9 9 9 9 E + 1 9 ,

g

ID number Data number

Date

Time

Weighing data

Advertising