Manual Response Adjustment


1 Press and hold the

MODE

key until

RESPONSE

is

displayed, then release the key.
Press the

MODE

key again quickly.





2 Select a stage of the response adjustment using

the

MODE

key. Either

FAST

,

MID.

or

SLOW

can be selected.




3 The balance displays

1

end

1

, returns to the

weighing mode and displays the updated
response indicator for about thirty seconds.




Advise

If the automatic response adjustment is not helpful, specify a parameter for "Condition (

Cond

)

" of "Environment, Display (

ba5fnc

)" with key operation.

Release and wait

Press and hold

Release

Press again

immediately

Select a parameter