Storing the calibration and calibration test data – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 58

Advertising
background image

56

12.3.

Data Memory for Calibration and Calibration Test

Calibration

data

(when

and how it is performed) and calibration test

data can be stored in memory.

All the data in memory is available to be output at one time to a

printer or personal computer.

Up to 50 data sets of the latest calibration or calibration test can be

stored. When the memory capacity has been reached, "

fUl

"

illuminates in the upper left of the display as shown below.


Storing the calibration and calibration test data

Note If "

Clr

" appears blinking in the upper left of the display, delete the data in memory.

Store the calibration and calibration test data as follows:

1 Set the "Data memory (

data

)" parameter to "

2

".

2 Set the "GLP output (

info

)" parameter to "

1

" or "

2

".

3 With the settings above, each time calibration or calibration test is performed,

the data is stored automatically.


Transmitting the memory data

Note

Confirm that the "Serial interface (

5if

)" parameters are set properly.

Refer to "10. Function Table" and "15.2. Connection To Peripheral Equipment".

Confirm that the "Data memory (

data

)" parameter is set to "

2

".

1 Press and hold the

PRINT

key until

out

is displayed, then

release the key.

2 Press

the

PRINT

key to display

out "n"o

with "no" blinking.

3 Press

the

RE-ZERO

key to display

out "g"o

with "go" blinking.

4 Press

the

PRINT

key to transmit all memory data using the RS-232C interface.

5 The balance displays

Clear

when all memory data is output.

Press the

CAL

key to return to the weighing mode.


Deleting data stored in memory

1 Press and hold the

PRINT

key until

out

is displayed, then release the key.

2 Press

the

SAMPLE

key to display

Clear

.

3 Press

the

PRINT

key to display

Clr "n"o

with "no" blinking.

4 Press

the

RE-ZERO

key to display

Clr "g"o

with "go" blinking.

5 Press

the

PRINT

key to delete all data.

6 The balance displays

out

and returns to the weighing mode

when all the data has been deleted.