Commands, Command list – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 66

Advertising
background image

64

15.3.

Commands

15.3.1.

Command List

Note A command has a terminator added, that is specified using "

5if Crlf

" of the

function table, and is sent to the balance.

Commands to query weighing data
C

Cancels the S or SIR command.

Q

Requests the weighing data immediately.

S

Requests the weighing data when stabilized.

SI

Requests the weighing data immediately.

SIR

Requests the weighing data continuously.

Commands to control the balance

CAL

Same as the

1

CAL

1

key.

OFF

Turns the display off.

ON

Turns the display on.

P

Same as the

ON:OFF

key

PRT

Same as the

PRINT

key

R

Same as the

RE-ZERO

key

(Note 1)

RNG

Same as the

RANGE

key.

TR

Tares the balance (Cancels the container’s weight.)

(Note 1)

TST

Perform calibration test.

U

Same as the

MODE

key

Commands to request data
?ID

Requests the identification number.

?PT

Outputs the tare value.

(Note 1)

?SN

Request the serial number of the balance.

?TN

Request the model name of the balance.

Command to set data
PT:∗∗∗.∗∗∗∗ g Sets the tare value.

(Note 1)

(Note 1)

R command assumes the point as zero and sets the display to zero.

TR command cancels the tare weight when the displayed value is greater than zero.

The canceled tare value can be read using the ?PT command.

PT: command sets the tare value digitally.

An example of this command sets a negative target value and fills with the sample
until the display becomes zero.

Step 1 Place a container.
Step 2 Set the display to zero using the R command.
Step 3 Set a target value using the PT: command.

For details, refer to page 69.


Advertising