Entering the density of a liquid

Two ways to set the density of a liquid are available in the function table, "Liquid density input
(

ldin

): by entering the water temperature or by entering the density directly.


1 Press

the

MODE

key as necessary to select the density mode. When

the density mode begins, "g (gram)" is displayed and the processing
indicator ( ) blinks.

Entering the water temperature (

ldin 0

)

2 The water temperature currently set (unit:°C, factory setting : 25°C) is

displayed. Use the following keys to change the value.

RE-ZERO

(+)key .. The key to increase the temperature by one degree.

(0°C is displayed after 99°C)

MODE

(-)key........ The key to decrease the temperature by one degree.

(99°C is displayed after 0°C)

PRINT

key ......... The key to store new water temperature, display

end

and

return to the density mode. Proceed to Step 1.

CAL

key............. The key to cancel the change and return to the density mode.

Proceed to Step 1.

The relation between the water temperature and density

°C

+0 +1 +2 +3 +4 +5 +6 +7 +8 +9

0 0.99984 0.99990 0.99994 0.99996 0.99997 0.99996 0.99994 0.99990 0.99985 0.99978

10 0.99970 0.99961 0.99949 0.99938 0.99924 0.99910 0.99894 0.99877 0.99860 0.99841
20 0.99820 0.99799 0.99777 0.99754 0.99730 0.99704 0.99678 0.99651 0.99623 0.99594
30 0.99565 0.99534 0.99503 0.99470 0.99437 0.99403 0.99368 0.99333 0.99297 0.99259
40 0.99222 0.99183 0.99144 0.99104 0.99063 0.99021 0.98979 0.98936 0.98893 0.98849
50 0.98804 0.98758 0.98712 0.98665 0.98618 0.98570 0.98521 0.98471 0.98422 0.98371
60 0.98320 0.98268 0.98216 0.98163 0.98110 0.98055 0.98001 0.97946 0.97890 0.97834
70 0.97777 0.97720 0.97662 0.97603 0.97544 0.97485 0.97425 0.97364 0.97303 0.97242
80 0.97180 0.97117 0.97054 0.96991 0.96927 0.96862 0.96797 0.96731 0.96665 0.96600
90 0.96532 0.96465 0.96397 0.96328 0.96259 0.96190 0.96120 0.96050 0.95979 0.95906


Entering the density directly (

ldin 1

)

The density currently set (unit : g / cm

3

, factory setting : 1.0000g / cm

3

) is displayed.

Use the following keys to change the value.

The range to set the density is 0.0000g / cm

3

to 1.9999g / cm

3

.

RE-ZERO

1

(+)key.. The key to set the value of the digit selected.

RANGE

1

(-)key...... The key to select the digit to change the value.

PRINT

key ......... The key to store the change, display

end

and return to the density

mode. Proceed to Step 1.

CAL

key............. The key to cancel the change and return to the density mode. Proceed to

Step 1.