Initializing the balance – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 32

Advertising
background image

30

9.2.

Initializing the Balance

This function returns the following parameters to factory settings.

Calibration

data

Function

table

The sample unit mass value (counting mode),

100% reference mass value (percent mode)

The data that is stored in the balance using the data memory function
External calibration weight and target mass value
Function switch settings ("9.1. Permit Or Inhibit")
Liquid density and temperature in the density mode


Note Be sure to calibrate the balance after initialization.


1 Press

the

ON:OFF

key to turn off the display.



2 While pressing and holding the

PRINT

key and the

RANGE

key, press the

ON:OFF

key to display

p5

.



3 Press

the

RANGE

key to display

Clr

.



4 Press

the

PRINT

key.

To cancel this operation, press the

CAL

key.



5 Press

the

RE-ZERO

key to display

Clr go

.



6 Press

the

PRINT

key to initialize the balance.

The balance will automatically return to the weighing mode.





Press and hold Press

Advertising