Stability band width ( 5t-b ), Zero tracking ( trc ), Display refresh rate ( 5pd ) – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 39: Decimal point ( pnt ), Auto display-on ( p-on )

Advertising
background image

37

10.3.

Description of the Class "Environment, Display"

Condition ( Cond )

Cond 0

This parameter is for sensitive response to the fluctuation of a mass value. Used
for powder target mass, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays

FAST

.

Cond 2

This parameter is for stable weighing with slow response. Used to prevent a mass
value from drifting due to vibration or drafts. After setting, the balance displays

SLOW

.

Notes

In automatic response adjustment, this parameter is selected automatically.

Stability band width ( 5t-b )

This item controls the width to regard a mass value as a stable value. When the

fluctuation per second is less than the parameter, the balance displays the stabilization
indicator and outputs or stores the data. The parameter influences the "Auto print mode"

5t-b

0 This parameter is used for sensitive response of the stabilization indicator. Used

for exact weighing.

5t-b

2 This parameter ignores slight fluctuations of a mass value. Used to prevent a mass

value from drifting due to vibration or drafts.

Zero tracking ( trc )

This function tracks zero point drift caused by changes in the environment and

stabilizes the zero point. When the weighing data is only a few digits, turn the function
off for accurate weighing.

trc 0

The tracking function is not used. Used for weighing a very light sample.

trc 1

The normal tracking function is used.

trc 2

The strong tracking function is used.

trc 3

The very strong tracking function is used. Used for stable zero display.

Display refresh rate ( 5pd )

The period to refresh the display. This parameter influences "Baud rate", "Data output

pause" and the data output rate of "Stream mode".

Decimal point ( pnt )

The decimal point format can be selected.

Auto display-ON ( p-on )

When the AC adapter is plugged in, the display is automatically turned on without the

1

ON:OFF

key operation, to display the weighing mode. Used when the balance is built

into an automated system. one hour warm up is necessary for accurate weighing.

Advertising