16. Maintenance

16.1.

Treatment of the Balance

Clean the balance with a lint free cloth that is moistened with warm water and a mild

detergent.

Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs

service or repair.

Use the original packing material for transportation.
Consider "3. Precautions" when operation the balance.



16.2.

Error Codes

Display Error

code

Description

EC,E11

Stability error
The balance can not stabilize due to an environmental
problem. Check around the pan. Prevent vibration, drafts,
temperature changes, static electricity and magnetic fields,
from influencing the balance.
To return to the weighing mode, press the

1

CAL

1

key.

EC,E16

Internal mass error
Applying the internal mass does not yield a change in the
mass value as specified.
Confirm that there is nothing on the pan and perform the
weighing operation from the beginning again.

EC,E17

Internal mass error
The internal mass application mechanism does not function
properly. Perform the weighing operation from the beginning
again.

EC,E20

Calibration weight error
The calibration weight is too heavy. Confirm the calibration
mass value. Press the

1

CAL

1

key to return to the weighing

mode.

EC,E21

Calibration weight error
The calibration weight is too light. Confirm the calibration
mass value. Press the

1

CAL

1

key to return to the weighing

mode.

Overload error
A sample beyond the balance weighing capacity has been
placed on the pan. Remove the sample from the pan.

Weighing pan Error
The mass value is too light.
Confirm that the weighing pan is properly installed and
calibrate the balance.