8.4.

Calibration Test Using the Internal Mass

This function tests the balance accuracy using the internal mass. (Balance is not calibrated)
When the GLP output is set, the calibration test report is output or stored.


1 Connect the AC adapter and warm up the

balance at least one hour.


2 Press and hold the

CAL

key until is

CC in

displayed, then release the key.



3 The balance measures the zero point.

Prevent vibration and drafts to affect the balance.



4 The measured zero point data is displayed.


5 The balance measures the full scale data.

Prevent vibration and drafts to affect the balance.


7 The full scale data is displayed. The tolerance

of the full scale data is ±0.2mg.

Model

Full scale data

GH-120 100.000

g

GH-200, GH-300,
GH-202, GH-252

200.000 g



8 When the GLP output is set, the calibration

test report is output or stored.

Refer to "GLP output (

info

)" and "Data

memory (

data

)" of the function table.

The zero point data and full scale data is

displayed (or output) in unit of 0.0001g.





8 The balance automatically returns to the weighing mode.

Press and hold

Release