A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 64

Advertising
background image

62

1 2

3 4

MODE

ON

1 2

3 4

MODE

ON

15.2.

Connection to Peripheral Equipment

Connection to an AD-8121B printer

Preset the following parameters to use the AD-8121B printer.

Class

Item and Parameter Factory

settings

AD-8121B

MODE 1

AD-8121B

MODE 2

AD-8121B

MODE 3

prt

Data output mode

0

0,1,2

3

0,1,2

ap-p

Auto print polarity

0

ap-b

Auto print difference

1

#1

Not

necessary

#1

d-no

Data number output

0

0

0

0,1

5-td

Time/Date output

0

0

0

0,1,2,3

5-id

ID number output

0

0

0

0,1

pU5e

Data output pause

0

0

0

0,1 #2

dout
Data output

at-f

Auto feed

0

0

0

0,1

bp5

Baud rate

2

2

2

2

btpr

Data bit, parity bit

0

0

0

0

Crlf

Terminator

0

0

0

0

type

Data format

0

0

0

1

5if
Serial
interface

Ct5

CTS, RTS control

0

0

0

0

#1 Set parameters when auto print mode A or B (prt 1 or 2) is selected.
#2 Set 1 when multiple lines are printed. Example: When appending ID number, set 1.


Notes

Refer to "11.2. GLP Report" concerning print samples.
Settings of AD-8121B

MODE

AD-8121B

DIP switch

Description

MODE 1

1 2

3 4

MODE

ON

Print at receiving data.

Standard mode, statistic mode

MODE 2

1 2

3 4

MODE

ON

Print by DATA key operation or built-in timer.

Standard mode, interval mode, chart mode

MODE 3

1 2

3 4

MODE

ON

Print at receiving data.

Dump print mode

DIP switch No.3 : Handling unstable data

ON

Print

OFF

Not

printed

DIP switch No.4 : Data input specifications (Interface selection)

ON

Current

loop

OFF

RS-232C

Advertising