Other display – A&D A & D Analytical Balance GH-120/GH-200/GH-300/GH-202/GH-252 User Manual

Page 74

Advertising
background image

72


Display Error

code

Description

EC,E00

Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.

EC,E01

Undefined command error
An undefined command was received.
Confirm the command.

EC,E02

Not ready
A received command can not be processed.
Example:

The balance received a "Q" command, but not in the

weighing mode.

The balance received a "Q" command while processing a

RE-ZERO command. Adjust the delay time to transmit a
command.

EC,E03

Timeout error
If the timeout parameter is set to "

t-Up1

", the balance did not

receive the next character of a command within the time limit
of one second. Confirm the communication.

EC,E04

Excess characters error
The balance received excessive characters in a command.
Confirm the command.

EC,E06

Format error
A command includes incorrect data.
Example:

The data is numerically incorrect.

Confirm the command.

EC,E07

Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.

Other error code

If an error described above can not be released or other
errors are displayed, contact the local A&D dealer.


16.3.

Other Display

When this indicator blinks, automatic self calibration is required. The
indicator blinks when the balance detects changes in ambient
temperature. If the balance is not used for several minutes with this
indicator blinking, the balance performs automatic self calibration. The
blinking period depends on the operating environment.

Advise The balance can be used while this indicator is blinking. We

recommend that you perform automatic self calibration for precision
weighing.

Advertising