8.5.

Calibration Using an External Weight

This function calibrates the balance using an external weight.


1 Connect the AC adapter and warm up the balance for at

least one hour with nothing on the weighing pan.


2 Press and hold the

CAL

key until

Calout

is

displayed, then release the key.


3 The balance displays

Cal 0

.

If you want to change the calibration mass, press

the

RANGE

key and proceed to step 4.

If you use the calibration mass value stored in the

balance, proceed to step 5.


4 Specify the calibration mass value as follows:

RANGE

key........ The key to switch blinking figures.

RE-ZERO

(+)key ... The keys to select the calibration mass or

adjust the mass value. Refer to page 23.

PRINT

key.......... The key to store the new mass value. Even

if the AC adapter is removed, the data is
maintained in non-volatile memory.

CAL

key............. The key to cancel the operation and return

to

Cal 0

.




5 Confirm that there is nothing on the pan and press the

PRINT

key. The balance measures the zero point.

Do not allow vibration or drafts to affect the balance.

The balance displays the calibration mass value.


6 Place the displayed calibration weight on the pan and press

the

PRINT

key. The balance measures the calibration

mass. Do not allow vibration or drafts to affect the balance.


7 The balance displays

end

. Remove the weight from the pan.


8 If the "GLP output (

info

)" the function table is set, the

balance displays

glp

and outputs or stores "Calibration

Report". Refer to "11.2. GLP Report" for details.


9 The balance will automatically return to the weighing mode.


10 Place the calibration weight on the pan and confirm that the

value displayed is correct. If it is not within the range, check
the ambient conditions such as breeze and vibration also
check the weighing pan. Then, repeat steps 1 to 10.

Release

Select

Example:

New weight

100.0012 g

Place weight

Remove

GLP output

Press

and

hold

Specify

MODE