AX-T-314A-S External key Plug

This plug produces the same operations as
pressing the

RE-ZERO

and

PRINT

keys. It

enables remote control of the balance using
an external switch.

This operation must connect the GND line to

the PRINT or RE-ZERO line for at least 100
miliseconds.





AX-SW128 Foot switch

The switch can be used to connect the
external key plug and used to the

RE-ZERO

or

PRINT

keys.

Note When the foot switch is to be used,

prepare the AX-SW128 and AX-T-314A-S
beforehand.



























Foot switch

AX-SW128

External key Plug

AX-T-314A-S

Terminal