10.2.

Details of the Function Table

Class Item

and

Parameter

Description

0

Fast response, sensitive value

FAST

1

MID.

Cond
Condition

2

Slow response, stable value

SLOW

Can be changed by response

adjustment.

0

Stable range is ±1 digit

1

#1

5t-b
Stability band width

2

Stable range is ±3 digits

The stabilization indicator illuminates

when the display fluctuation is within

the range.

0

OFF

1

Normal

2

Strong

trc
Zero tracking

3

Very strong

Keeps zero display by tracking zero

drift.

0

5 times/second

5pd
Display refresh rate 1

10 times/second

Period to refresh the display

0

Point (.)

pnt
Decimal point

1

Comma (,)

Decimal point format

0

OFF

ba5fnc
Environment
Display

p-on
Auto display-ON

1

ON

Turns on the weighing mode display

when the AC adapter is connected.

Cl adj
Clock

Refer to "10.7. Clock and Calendar Function"

The time and date are added to the

output data.

0

Key mode

Accepts the PRINT key only when

the display is stable.

1

Auto print mode A
(Reference = zero)

2

Auto print mode B
(Reference = last stable value)

Outputs data when the display is

stable and conditions of ap-p, ap-b

and the reference value are met.

prt
Data output mode

3

Stream mode /
Interval memory mode

With data 0, outputs data

continuously; with data 2, uses

interval memory.

0

Plus only

Displayed value>Reference

1

Minus only

Displayed value<Reference

ap-p
Auto print polarity

2

Both

Regardless of displayed value

0

10 digits

1

100 digits

ap-b
Auto print difference

2

1000 digits

Difference between reference value

and displayed value

#2

0

Not used

1

Stores weighing data

dout
Data output

data
Data memory

2

Stores calibration data

Related items:

prt, int, d-no, 5-td, info

: Factory settings.

#1 The unit of minimum display is digit.

Example: If 1 mg display is selected using the

RANGE

key for the GH-300, 1 mg is one

digit.

#2 Usable minimum display of the balance is one digit.

Example: In gram display, one digit is 0.01 mg for the GH-252 and 0.1 mg for the GH-300.