Myron L 9PTK User Manual

Page 14

Advertising
background image

10

4. pH and ORP/Fr Chl Keys

Measurements are made on solution held in the pH/ORP sensor well

(ref. pH and ORP/Free Chlorine, pg. 62). The protective cap is removed

and the sensor well is filled and rinsed with the sample enough times

to completely replace the storage solution.

After use, the pH/ORP sensor well must be refilled with Myron L

Storage Solution, and the protective cap reinstalled securely (ref.

Maintenance of the pH/ORP Sensor, pg. 11 and Cleaning Sensors, 2.

pH/ORP, pg. 50).

a. pH Key

A press of

displays pH readings. No units are displayed on the

right.

b. ORP Key

A press of

displays Oxidation-Reduction Potential/REDOX

reading in millivolts if the unit is in ORP mode, or Free Chlorine

reading in parts per million if the unit is in Free Chlorine mode.

5. CAL/MCLR Key

While measuring conductivity, TDS, or pH, a press of

allows

you to enter the calibration mode. Once in CAL mode, a press of this

key accepts the new value. If no more calibration options follow, the

instrument returns to measuring (ref. Leaving Calibration, pg. 27).

If

is held down for about 3 seconds, CAL mode is not entered,

but “sel” appears to allow Solution Selection (ref. pg. 22) with the

Up or Down keys. As in calibration, the cal key is now an “accept”

key.

While measuring ORP or Free Chlorine, holding cal down for about

3 seconds allows ORP mode selection (ref. pg. 21).

For titrations and LSI Calculator, the cal key steps you through the

procedure and accepts values for final calculations.

Once in cal mode, a press of this key accepts the new value.

While reviewing stored records, the mclr side of the key is active

to allow clearing records (ref. Clearing a Record/Memory Clear, pg.

37).

6. UP or DOWN Keys

While measuring in any parameter, the

or

keys activate

Advertising