Myron L 9PTK User Manual

Page 19

Advertising
background image

15

3.

Rinse the conductivity cell cup and cell extender 3 times with

solution to be measured.

4.

Refill cell cup and cell extender with sample solution.

5.

Insert T-plunger into open cell extender until the “T” rests flush

on the lip of the cell extender. This will cause fluid beyond what

is required for titration to overflow the cell extender.

6.

Rotate the T-plunger from side to side, allowing the arms of the

T-plunger to slide along the rim of the cell extender.

7.

Carefully remove the T-plunger so as not to flick or spill remaining

sample. Tap the T-plunger on the side of the cell extender to

remove sample that may be stuck to the T-plunger.

8.

Press

. “al alternating with “tds” briefly displays along

with the “PPm” value of the sample. “P

r

es cal” will display

when the reading stabilizes.

9.

Press cal to accept and advance to the first titration prompt.

add” alternating with “a1” displays; then “agit t1” alternating

with “hOld” displays. (The number after “t” indicates which

titration point you are on. The number “1” here indicates this is

the first titration point. Subsequent titration points are sequenced

numerically: 2, 3, 4, etc.)

10.

Using the pipette, add 100 µL of Reagent: A1, close the cell

extender cap securely, then agitate and hold. Repeat as

prompted. “P

r

es cal” will display.

11.

Press cal to advance to the next titration point. “add

alternating with “a1” displays; then “agit t2” alternating with

hOld” displays.

12.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

13.

Continue adding 100µL of Reagent A1, agitating and holding the

Advertising