Myron L 9PTK User Manual

Page 35

Advertising
background image

31

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

Hardness Standard 200PPM.

4.

Refill cell cup and cell extender with Hardness Standard 200PPM.

5.

Insert T-plunger into open cell extender until the “T” rests flush

on the lip of the cell extender. This will cause fluid beyond what

is required for titration to overflow the cell extender.

6.

Rotate the T-plunger from side to side, allowing the arms of the

T-plunger to slide along the rim of the cell extender.

7.

Carefully remove the T-plunger so as not to flick or spill remaining

sample. Tap the T-plunger on the side of the cell extender to

remove sample that may be stuck to the T-plunger.

8.

Press

. “ca” alternating with “tds” displays along with

PPm” sample value. “P

r

es cal” displays when the reading

stabilizes.

9.

Press cal. “add” alternating with “c1” will display; then “agit

alternating with “hOld” will display.

10.

Using the pipette, add 100µL of Reagent C1 to the sample, close

the cell extender cap securely, then agitate and hold. Repeat as

prompted. “OPe

n

caP” alternating with “P

r

es cal” will display.

11.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

12.

Press cal to start 120 second timer. The display will count

down from “120” to “0”, allowing time for CO

2

to escape. Using

your index finger, firmly tap the side of the instrument closest