Myron L 9PTK User Manual

Page 38

Advertising
background image

34

5.

Press

or

until the display reads “7.00”.

nOte: Attempted calibration of >1 pH point from factory calibration will

cause “Fac” to appear. This indicates the need for sensor replacement

(ref. Troubleshooting pg. 52) or fresh buffer solution. The “FAC” internal

electronic calibration is not intended to replace calibration with pH

buffers. It assumes an ideal pH sensor Each “FAC” indicates a factory

setting for that calibration step (i.e., 7, acid, base).

You may press

to accept the preset factory value, or you may

reduce your variation from factory setting by pressing

or

.

6.

Press

to accept the new value. The pH Zero Calibration

is now complete. You may continue with pH gain Calibration or

exit by pressing any measurement key.

b. pH gain Calibration

imPOrtant: Always calibrate or verify your Ultrameter III with a pH

7 buffer solution before adjusting the gain with acid or base buffers,

i.e., 4 and/or 10, etc. Either acid or base solution can be used for the

2nd point “gain” calibration and then the opposite for the 3rd point. The

display will verify that a buffer is in the sensor well by displaying either

acd” or “bas”.

1.

The pH calibration mode is initiated by either completion of the

pH Zero Calibration, or verifying 7 buffer and pressing the

key twice while in pH measurement mode.

2.

At this point the “cal”, “bUFFer” and “acd” or “bas

annunciators will be displayed (see Figures 7 and 8).

Figure 8

pH

BUFFER

CAL

nOte: If the “acd” and “bas” indicators are blinking, the unit is indicating an

error and needs either an acid or base solution present in the sensor well.

Advertising