Agitate and hold as many times as prompted.

Allow the display to cycle through all screens, e.g.,

edta”, “agit” and “hOld”, at least once. Continue

holding steady until “P

r

es cal” displays.

Hold the unit steady before pressing cal to

accept readings.

C. Measuring Conductivity & Total Dissolved Solids (TDS)

1.

Rinse cell cup 3 times with sample to be measured. (This

conditions the temperature compensation network and prepares

the cell.)

2.

Refill cell cup with sample.

3.

Press

or

.

4.

Take reading. A display of [- - - -] indicates an overrange

condition.

D. Measuring Resistivity

Resistivity is for low conductivity solutions. In a cell cup the value may

drift from trace contaminants or absorption from atmospheric gasses, so

measuring a flowing sample is recommended.
1.

Ensure pH protective cap is secure to avoid contamination.

2.

Hold instrument at 30° angle (cup sloping downward).

3.

Let sample flow continuously into conductivity cell with no aeration.

4.

Press

key; use best reading.

nOte: If reading is lower than 10 kilohms, display will be dashes:

[ - - - - ]. Measure solution Conductivity instead.

E. Measuring Alkalinity

Prepare the materials required for titration: Reagent: A1, cell extender,

T-plunger, and 100 µL pipette with a clean tip installed.
nOte: If you do not complete each titration point within the 3-minute

timeout period, the unit exits the Alkalinity function. Between titrations,

the unit will automatically power off after the period of inactivity defined

in the AUTO OFF setting (ref. AUTO OFF, pg. 43). Press alK to start the

titration process over from step 1.
If you make a gross error in your titration method, e.g., you forget to add

sample to the cell, “err” will display. Start the titration over from Step 1

to obtain a reading.
1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

hOld”: When “hOld

displays, hold the 9P

still with bottom of case

parallel to the ground.