User mode calibratiOn linc™ FUnctiOn

The Linc™ function allows easy calibration when in User mode and the

user does not have a user standard solution to calibrate the instrument.

This function will ensure more repeatable and accurate measurements

than many other calibration methods. It is recommended that this

function be used to provide the highest degree of confidence when

the Ultrameter III is used in User mode. When Linc is used, the User

mode is linked to another standard, i.e., if User and KCl are linked,

a KCI standard solution is used to calibrate the instrument. It is that

simple.

A. Calibration of Ultrameter III for use in User mode

1.

Press

or

key.

2.

Calibrate the unit using a Standard Solution (ref.

CALIBRATION, pg. 26).

3.

Place the Ultrameter III in User mode (ref. SOLUTION

SELECTION, pg. 22).

4.

Verify/Set the calibration linc. (See below – Setting “User”

Mode Calibration “Linc”.)

B. Setting User mode Calibration “Linc”

The Linc function sets or “links” the calibration gain factor of a Standard

Solution to the User solution mode. Once set, the “linc” will stay intact

with future calibrations unless the Linc has been canceled. For more

information on canceling the User mode Calibration Linc refer to the

section “Canceling User mode Calibration “linc””, pg. 46.