Myron L 9PTK User Manual

Page 17

Advertising
background image

13

conductivity cell being careful not to

touch the tip to the existing solution.

Depress the top of the pipette to the

SECOND STOP (all the way down)

being careful to keep the tip over the

cell. Release the plunger button and let

it return to the REST position.

2. T-Plunger Instructions

To PLUNgE the cell:

With the cell extender installed and

solution in the cell, insert the tip of the

T-plunger in the cell extender until the

arms of the T-plunger are flush against

the rim of the cell extender. Solution will

overflow the cell.
Keeping the arms of the T-plunger flush

against the rim of the cell extender, rotate

the T-plunger from side to side, allowing

the arms of the T-plunger to slide along

the rim of the cell extender.
Remove the T-plunger from the cell by

lifting it straight up out of the cell, being

careful to keep it centered over the

mouth of the cell extender. When the

T-plunger is completely out of solution,

tap the T-plunger once firmly on the rim

of the cell extender to dispel any clinging

solution back into the cell.
Remove the T-plunger completely from

the cell area.

3. Mixing Solution in Cell

a. Agitate

When the display says to agitate (“agit”): grasp

the instrument on both sides of the keypad (keypad

facing up) with your hands so you don’t accidentally

drop or release it. Agitate the solution in the cell by

swinging the cell cup area of the unit up and down 3

times. Do NOT shake the instrument vigorously as

this could create bubbles in the sample solution.

b. Hold

When the display says to hold (“hOld”):

Hold the instrument steady in the horizontal position

while the instrument records a measurement.

Insert the T-plunger all the way

down into the cell extender then

rotate it from side to side.

Hold the T-plunger over the cell

then tap firmly so that solution is

dispelled back into the cell cup.

agit”: When “agit

displays, swing the

cell end of the 9P up

and down 3 times.