Myron L 9PTK User Manual

Page 55

Advertising
background image

51

Then use one of the following methods:

1.

Pour a HOT salt solution ~60°C/140°F — a potassium

chloride (KCI) solution such as Myron L pH/ORP Sensor

Storage Solution is preferable, but HOT tap water with table

salt (NaCl) will work fine — in the sensor well and allow to

cool. Retest.

or

2.

Pour DI water in the sensor well and allow to stand for no more

than 4 hours (longer can deplete the reference solution and

damage the glass bulb). Retest.

If neither method is successful, the sensor must be replaced.

“Drifting” can be caused by a film on the pH sensor bulb and/or reference.

Use isopropyl alcohol (IPA) or spray a liquid cleaner such as Windex™

or Fantastic™ into the sensor well to clean it. The sensor bulb is very

thin and delicate. Do not scrub or wipe the pH/ORP sensor.

Leaving high pH (alkaline) solutions in contact with the pH sensor for

long periods of time is harmful and will cause damage. Rinse such

liquids from the pH/ORP sensor well and refill it with Myron L Storage

Solution to extend the useful life of the sensor. If unavailable, you can

use a saturated KCl solution, pH 4 buffer, or a saturated solution of

table salt and tap water.

Samples containing chlorine, sulfur, or ammonia can “poison” any pH

electrode. If it is necessary to measure the pH of any such sample,

thoroughly rinse the sensor well with clean water immediately after

taking the measurement. Any sample element that reduces (adds

an electron to) silver, such as cyanide, will attack the reference

electrode.

Replacement sensors are available only from the Myron L Company

or its authorized distributors (ref. Replacement pH/ORP Sensor, pg.

56).

it is possible to desensitize the OrP sensor when measuring

high OrP solutions. to check the OrP sensor functioning, take

an OrP reading of myron l ph/OrP sensor storage solution. if

the reading is outside the range of 350-400 mv, clean Only the

platinum OrP electrode with myron l OrP conditioner solution-

soaked cotton swab, being careful not to touch the swab to the

glass bulb of the ph sensor.