Page 26

Advertising
background image

22

2.

Remove protective cap by rotating while grasping and pulling up.

3.

Rinse sensor well and cell cup 3 times with sample to be

measured. Shake out each sample to remove any residual liquid.

4.

Refill both sensor well and cell cup with sample.

5.

Press

.

6.

Take reading.

7.

Press ms to store reading in memory, if desired.

imPOrtant: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap. If

Myron L pH Sensor Storage Solution is unavailable, you can

use a strong KCl solution, a pH 4 buffer, or a saturated solution

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 50). Do not allow pH/ORP sensor to dry out.

3. Measuring Free Chlorine

1.

Ensure the 9P is in free chlorine mode (ref. ORP Mode

Selection, pg. 21).

2.

Remove protective cap by rotating while grasping and pulling up.

3.

Thoroughly flush the sensor well and cell cup with a steady

stream of the solution you intend to measure by allowing the

solution to flow into and out of the sensor well and cell cup for

at least 10 seconds.

4.

Refill both sensor well and cell cup with sample.

5.

Press

.

6.

Take reading. Annunciators alert you when the concentration is

outside the specified measurement range. “-Or-” (over range)

will display when the concentration is over the range limit (>

10.00 ppm).

7.

Press ms to store reading in memory if desired.

imPOrtant: After use, fill pH/ORP sensor well with Myron L

pH Sensor Storage Solution and replace protective cap. If

Myron L pH Sensor Storage Solution is unavailable, you can

use a strong KCl solution, a pH 4 buffer, or a saturated solution

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 50). Do not allow pH/ORP sensor to dry out.

vi.

sOlUtiOn selectiOn

A. Why Solution Selection is Available

Conductivity, resistivity, and TDS require temperature correction to

25°C values (ref. Standardized to 25°C, pg. 56). Selection determines

the temperature correction of conductivity and calculation of TDS from