Myron L 9PTK User Manual

Page 56

Advertising
background image

52

Xviii. trOUbleshOOting chart

symptom

Possible cause

corrective action

No display, even though

measurement key pressed

Battery weak or not connected.

Check connections or replace battery. Ref. Battery Replacement, pg. 50.

Inaccurate ph readings

1. pH calibration needed. Ref. pH Cal.,

pg. 33.

2. Cross-contamination from residual pH

buffers or samples in sensor well.

3. Calibration with expired pH buffers.

1. Recalibrate instrument.

2. Thoroughly rinse sensor well.

3. Recalibrate using fresh buffers. Ref. pH Buffer Solutions, pg. 55.

No response to ph changes

Sensor bulb is cracked or an

electromechanical short caused by an

internal crack.

Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 56.

Will not adjust down to ph 7

pH/ORP sensor has lost KCl.

Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no

improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 56).

ph readings drift or respond slowly to

changes in buffers/samples or “Fac

is displayed repeatedly

1. Temporary condition due to memory” of

solution in pH sensor well for long periods.

2. Bulb dirty or dried out.

3. Reference junction clogged or coated.

Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no

improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 56).

Unstable conductivity/tDs/

resistivity readings

1. Dirty electrodes.

2. Actual resistance is changing due to

atmospheric contamination.

1. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.

2. Minimize test sample exposure to air by taking a flowing sample.

Ref. Measuring Resistivity, pg. 14.

Unable to calibrate conductivity/

tDs

Film or deposits on electrodes.

Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.

resistivity readings much lower than

expected

1. Contamination from previous sample or

from pH sensor well.

2. Carbon dioxide in test sample.

1. Rinse cell cup more thoroughly before measurement. Ensure pH cap is snugly in

place.

2. See Measuring Resistivity, pg. 14.

alkalinity or hardness titration with

calibration solution value not as

stated on bottle

1. Method error.

2. Dirty Cell Cup.

3. Dirty pipette.

4. Faulty pipette.

5. Contaminated calibration solution.

1. Repeat titration.

2. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.

3. Clean pipette per manufacturer’s instructions.

4. Use a new pipette.

5. Use fresh calibration solution. Be sure to use a new tip for each type of solution

measured.

Low Reading

Slow or no response to OrP changes

ORP platinum electrode is dirty.

Clean ORP electrode with ORP conditioner. Ref. Cleaning Sensors, 2. pH/ORP

Sensor, pg. 50.

Advertising