vii. calibratiOn

A. Calibration Intervals

generally, calibration is recommended about once per month with

Conductivity or TDS solutions. Calibration with pH solutions should

be checked twice a month. Calibration of ORP is not necessary (ref.

CALIBRATION INTERVALS TABLE under Calibration Steps below).

B. Rules for Calibration of the Ultrameter III

1. Calibration Steps

The following table lists measurement functions and their corresponding

calibration requirements:

CALIBRATION INTERVALS TABLE:

Function

Kcl, nacl or 442

User

Cond

gain Only

Tempco, then gain

Res

Done in Conductivity

Done in Conductivity

TDS

gain Only

Tempco, Ratio, then gain

pH

7, acid and/or base

ORP

Zero set with pH 7

automatically

Alkalinity

gain Only

Hardness

gain Only

a. Starting Calibration

For Alkalinity and Hardness, calibration is begun by pressing

when the titration is complete and a value displays.

For Conductivity, TDS or pH, calibration is begun by pressing

while measuring Conductivity, TDS or pH. Measuring continues, but

the “cal” icon is on, indicating calibration is now changeable. The

reading is changed with the

and

keys to match the

known value.

The calibration for each of the 4 solution types may be performed in

either conductivity or TDS mode.

Once in CAL mode, the

key becomes an “ACCEPT” key for the

new calibration.

For pH, at each point, pressing

accepts the new calibration

value and steps you to the next adjustment.